| Catalog Number |
010000850 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Codes |
Pain (1994); Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D6a - implant date: unknown month and day in 2024.D10: cat# 00625006525 lot# j7298651 bone scr 6.5x25 self-tap, unk avenir size 3 taper 12/14, unk cocr 32mm femoral head.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the patient underwent a hip procedure on an unknown date.Subsequently, the patient was revised due to pain and instability.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; d5; g3; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were provided and review of the available records identified the following: single ap view of the bilateral hips demonstrate right total hip arthroplasty with screw fixation of the acetabular cup.Normal acetabular inclination angle.No radiolucency.Question mild heterotopic ossification along the acetabular superiorly.A definitive root cause cannot be determined.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information at the time of this report.
|
| |
|
Search Alerts/Recalls
|