Catalog Number 00625006525
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Date of Event 06/02/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient underwent a hip procedure on an unknown date.Subsequently, the patient was revised due to pain and instability.Attempts have been made, and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D6a - implant date: unknown month and day in 2024.D10: cat# 110010245 lot# 66226252 g7 osseoti 4 hole shell 54mm f, cat# 010000850 lot# 7156059 g7 neutral e1 liner 32mm f.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; d5; g3; h2; h3; h4; h6; h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were reviewed and identified the following: single ap view of the bilateral hips demonstrate right total hip arthroplasty with screw fixation of the acetabular cup.Normal acetabular inclination angle.No radiolucency.Question mild heterotopic ossification along the acetabular superiorly.A definitive root cause cannot be determined.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 25 MM LENGTH
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22308511
Report Number0001822565-2025-02067
Device Sequence Number12743822
Product Code LPH
UDI-Device Identifier00889024119819
UDI-Public(01)00889024119819(17)320708(10)J7298651
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K934765
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00625006525
Device Lot NumberJ7298651
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date06/23/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-