| Model Number |
VTICM5 12.6 (-8.00/1.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/31/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Lens was exchanged with a longer length lens.Problem resolved.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned in microcentrifuge vial with moisture.Visual inspection found haptic bent.Dimensional inspection found the lens to be within specification.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|