| Model Number |
9849 |
| Medical Device Problem Codes |
Inappropriate or Unexpected Reset (2959); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the rad-g was "getting on/off automatically." no patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: masimo has reached out to the customer to request the return of the device.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.E1.Initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).Health effect impact code: no adverse event, no patient impact.
|
| |
|
Event or Problem Description
|
|
The customer reported the rad-g was "getting on/off automatically." no patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned rad-g was evaluated.Shorting inside the on/off button was creating an electrical short across the button and resulting in the power button being activated even when not physically pressed.E1: initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).Masimo initiated a recall for this issue and notified us fda on 02/15/2024.The recall number is z-1538-2024.
|
| |
|
Search Alerts/Recalls
|