The reported event was unconfirmed because the reported failure could not be reproduced.The returned device was functionally tested and the product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.However, there was cid# (b)(4) opened for documented the investigation.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Instructions for use were reviewed for this investigation and found adequate.Bard®biocath® foley catheter (with lubricant).Single use only.Warnings: 1.Method for use.(1) do not inflate the balloon in the urethra.The urethra may be injured.(2) do not pull the catheter hard.The bladder/urethra may be injured.2.Applicable patients.(1) patients with delirium who might pull out catheter when patient tugs at catheter un-consciously, the bladder and urethra may be damaged.Contraindications: 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).They may damage the device and may burst balloon.(4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.Catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.2.Applicable patient: (1) do not use in patients who are or have been allergic to natural rubber shape, configuration and principles bard® biocath® foley catheter (with lubricant) consists of a balloon catheter and water-soluble lubricant.Material: balloon catheter: natural rubber latex, shape, product numbers, sizes etc.Relevant to this package insert are shown in package labeling.1.Balloon catheter.2.Accessories: water-soluble lubricant; intended use & effect-efficacy: this device is an indwelling catheter in the bladder for urinary drainage.Directions for use: 1.Method of use.The device is intended for single use only and is not reusable.(1) cleanse the area around the external urethral meatus with the cotton balls immersed in the antiseptics.(2) lubricate the distal end of the catheter with water-soluble lubricant.(3) insert catheter into the urethral meatus and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a needle-less syringe, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.(4) pull catheter to seat the balloon at the level of the bladder neck and secure placement.(5) to deflate balloon and remove catheter, insert a luertip(needleless) syringe in the inflation valve to allow the water to drain spontaneously.After balloon deflation, withdraw the catheter while confirming that no resistance is encountered.2.Precautions for use (1) when resistance is encountered in inserting catheter, stop the procedure and remove the catheter.(2) when deflating balloon, do not aspirate with a syringe.(3) when inserting catheter with a stylet, confirm that the stylet has reached the tip of the catheter, and make sure to keep the stylet in place inside the catheter during insertion.(4) when inflating balloon with sterile water, use the sterile water of the prescribed capacity labeled on the valve.(5) no substance except sterile water should be used to inflate the balloon.(6) do not wipe catheter surface with organic solvents such as alcohol.(7) do not aspirate urine through drainage funnel wall.(8) since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.(9) when urinary flow cannot be noted, confirm that the catheter is neither occluded nor 3 precautions: 1.Precautions for use (exercise caution when using the device in the following patients) (1) exercise caution when using the device in patients with high urinary calcium levels as encrustation on the balloon surface, catheter occlusion or damage may occur.2.Important precautions.(1) when catheter is inadvertently removed, inspect the balloon and shaft of catheter for rupture, defect, etc.Before inserting a new catheter.(2) if any possibility of retained balloon fragments exists, consider removing them using a cystoscope.(3) when it is difficult to remove catheter by deflating the balloon, take appropriate measures according to the section ¿troubleshooting¿.Troubleshooting: when it is difficult to remove catheter by deflating the balloon (expressed as ¿removal-difficult case¿ hereinafter), take appropriate measures according to the following procedures.The following two methods are available for removal-difficult cases.A.Non-rupture method (sterile water is withdrawn without bursting the balloon.) b.Balloon-rupture method.With balloon-rupture method, fragments of the ruptured balloon may remain in the bladder.Therefore, try non-rupture method first.A.Non-rupture method.1) attach luer tip syringe to the inflation valve.Inject an additional amount of sterile water into the inflation lumen and pump the plunger.2)if situation wouldn't be improved with 1), sever the inflation funnel of valve.(fig.1) 3) if situation wouldn¿t be improved with 2), sever the catheter shaft while holding it with for-ceps so that the distal segment may not be drawn into the urethra.(fig.2) 4) if situation wouldn't be improved with3), insert a needle into the inflation lumen and pump the plunger.(fig.3) 5) if situation wouldn't be improved with 4), insert a fine steel wire through the inflation lumen of catheter.(fig.4) b.Balloon-rupture method: 1) inject 100-200ml/cc of saline solution warmed to body temperature into the bladder through the drainage lumen, and then inject a large amount of water or 10-15 ml/cc of mineral oil into the balloon through the inflation lumen with a needle to induce rupture.After rupture of the balloon, irrigate the bladder.(fig.5) mineral oil.2) if situation wouldn't be improved with 1), attempt following procedures.A) under the radioscopic observation, infuse a contrast medium into the bladder, and burst the balloon by suprapubic puncture of the bladder.(fig.6) b) in male patients, burst the balloon by puncture with a needle from the perineal (or suprapubic) region or through the rectum under ultrasonographic guidance.C)in female patients, burst the balloon by insertion of a needle along the urethra.(fig.8) 3.Malfunction and adverse events: (1) malfunction: -catheter kinking, damage, rupture, -difficulty or failure to remove the device, -occlusion of catheter inner lumens, -encrustation, -accidental removal of the device due to leakage of sterile water or balloon rupture, -device damage due to inappropriate use.(2) adverse events: -urinary-tract infection, -hemorrhage, hematuria, -allergy reaction to the device, -calculus formation, -edema, -pain, -discomfort, -injury of bladder or urethral, -urethritis, urinary incontinence, -retained balloon fragments.[storage method and expiration date] 1.Storage.Correction: d.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|