| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the air/water cylinder and the air/water tube of the colonovideoscope had foreign objects.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device was returned for device investigation.The device evaluation found the air/water cylinder and the air/water tube of the colonovideoscope had foreign objects.There was no report on the subject device being used in procedure after the suggested event was reviewed.The customer provided the cleaning disinfection and sterilization (cds) processes where the use of simethicone and petroleum/oil/silicone-based lubricants are non-water soluble and thus were not recommended for use by olympus.Olympus will continue to monitor complaints for this device.
|
| |
|