Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was received: 1 hip ¿ pa.(b)(6), y936 - sbmemt, 104tka, 103kgm, y427 ¿ ramdtu, rgmkce, 4 returning, interrupted vicryl was used on the dermal layer, but the subcuticular dehisced with the 3-0 monocryl.Additional information was requested and the following was received: one suture used of y936 used and sbmemt & 104tka & 103kgm are possible lot numbers.One suture used of y427 used and ramdtu & rgmkce are possible lot numbers.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.Date and name of index surgical procedure? the diagnosis and indication for the index surgical procedure? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? how was the suture tied (square knot or multiple knots one end)? what tissue dehisced? please describe the appearance of the suture during the second procedure.Did the suture break? did the sutures untie? did the suture pull out of the tissue? were there any patient stress factors that led to the suture untying, breaking or pulling out of the tissue? onset date/time of dehiscence? (# post op days) how was the dehiscence managed? please describe any surgical intervention required for the wound dehiscence including date and findings.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? what symptoms did the patient experience following the index surgical procedure? onset date? what is the hospital¿s post-op protocol/exercise regime? did the patient follow the post-op protocol/exercise regime? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? d4: full udi currently unavailable since the exact lot of the suspected device cannot be determined.Additional information: d4, h4 - the actual device batch number associated with this event is not known.The following are reported possible batch numbers: batch sbmemt, expiration date: 1/31/2027 date of mfg.: 2/11/2022.Batch 103kgm, expiration date: aug/31/2029 date of mfg.: sep/24/2024.Batch 104tka, expiration date: oct/31/2029 date of mfg.: nov/19/2024.In addition, a review of the manufacturing record evaluation for the possible batch number was performed for the finished device lot, and no non-conformances were identified.H6 component code: g07002 no device problem.H3 investigation summary: the product was returned to ethicon for evaluation.One unopened sample was received for analysis.Product code y936, lot sbmem.In order to evaluate the condition of the returned sample, the packet was examined for visual defects: appearance, color package, wrinkles in the seal area, continuous seals, seal margins, over-sealing, damage (torn, punctured) and none was observed.The packet was opened, and the swage and attachment area were noted to be as expected.During the visual inspection, the suture was correctly placed on the winding former.The suture was dispensed without a problem and examined along of the strand and no anomalies or damaged on the suture surface could be observed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reach on the cause of the reported event.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance., the product was returned to ethicon for evaluation.One unopened sample was received for analysis.Product code y936, lot 104tka.In order to evaluate the condition of the returned sample, the packet was examined for visual defects: appearance, color package, wrinkles in the seal area, continuous seals, seal margins, over-sealing, damage (torn, punctured) and none was observed.The packet was opened, and the swage and attachment area were noted to be as expected.During the visual inspection, the suture was correctly placed on the winding former.The suture was dispensed without a problem and examined along of the strand and no anomalies or damaged on the suture surface could be observed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reach on the cause of the reported event.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance., the product was returned to ethicon for evaluation.One unopened sample was received for analysis.Product code y936, lot 103kgm.In order to evaluate the condition of the returned sample, the packet was examined for visual defects: appearance, color package, wrinkles in the seal area, continuous seals, seal margins, over-sealing, damage (torn, punctured) and none was observed.The packet was opened, and the swage and attachment area were noted to be as expected.During the visual inspection, the suture was correctly placed on the winding former.The suture was dispensed without a problem and examined along of the strand and no anomalies or damaged on the suture surface could be observed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reach on the cause of the reported event.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|