BECTON, DICKINSON & CO. (BROKEN BOW) VACUTAINER CPT; LYMPHOCYTE SEPARATION MEDIUM Back to Search Results
Catalog Number 362761
Medical Device Problem Code Device Ingredient or Reagent Problem (2910)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported during use of bd vacutainer® cpt¿ cell preparation tube with sodium citrate, 3 samples exhibited poor barrier separation.There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
Investigation summary: bd received two photos for investigation.Visual evaluation of these photos showed the presence of poor barrier separation.A total of 60 retained samples were visually inspected, with no issues being identified.Complaints for sample quality are under statistical control for the month of may 2025.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: sample quality.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
 
Event or Problem Description
It was reported during use of bd vacutainer® cpt¿ cell preparation tube with sodium citrate, 3 samples exhibited poor barrier separation.There was no health impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACUTAINER CPT
Common Device Name
LYMPHOCYTE SEPARATION MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key22311096
Report Number1917413-2025-00814
Device Sequence Number13008862
Product Code JCF
UDI-Device Identifier30382903627616
UDI-Public(01)30382903627616(11)240619(17)250630(10)4171848
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2025
Device Catalogue Number362761
Device Lot Number4171848
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2025
Supplement Date Received by Manufacturer06/30/2025
Initial Report FDA Received Date06/23/2025
Supplement Report FDA Received Date07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-