| Model Number |
3000TFX |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 9 years, 1 months due to stenosis.The procedure was performed successfully with a 26mm 9755rsl transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated sections: g3, g6, h6 investigation findings, and investigation conclusions.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia (hld), and diabetes mellitus (dm).
|
| |
|
Additional Manufacturer Narrative
|
|
Updated sections: b5, b7, g3, g6, h6 health effect, clinical code, and type of investigation.The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.
|
| |
|
Event or Problem Description
|
|
It was reported and through investigation that a patient with a 25mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 9 years, 1 months due to stenosis.The patient presented with hear failure and shortness of breath.The procedure was performed successfully with a 26mm 9755rsl transcatheter valve.The patient was in stable condition post procedure and was discharge on pod#1.
|
| |
|