• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES REPLACEMENT HEART-VALVE; CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES REPLACEMENT HEART-VALVE; CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS Back to Search Results
Model Number 3000TFX
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 05/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 25mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 9 years, 1 months due to stenosis.The procedure was performed successfully with a 26mm 9755rsl transcatheter valve.
 
Additional Manufacturer Narrative
Updated sections: g3, g6, h6 investigation findings, and investigation conclusions.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia (hld), and diabetes mellitus (dm).
 
Additional Manufacturer Narrative
Updated sections: b5, b7, g3, g6, h6 health effect, clinical code, and type of investigation.The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.
 
Event or Problem Description
It was reported and through investigation that a patient with a 25mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 9 years, 1 months due to stenosis.The patient presented with hear failure and shortness of breath.The procedure was performed successfully with a 26mm 9755rsl transcatheter valve.The patient was in stable condition post procedure and was discharge on pod#1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPLACEMENT HEART-VALVE
Common Device Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22312115
Report Number2015691-2025-05110
Device Sequence Number8947868
Product Code DYE
UDI-Device Identifier00690103163667
UDI-Public(01)00690103163667(17)190224
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/24/2019
Device Model Number3000TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2025
Supplement Date Received by Manufacturer07/03/2025
07/29/2025
Initial Report FDA Received Date06/23/2025
Supplement Report FDA Received Date07/28/2025
07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/25/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
Patient SexMale
-
-