|
EXACTECH, INC. OPTETRAK LOGIC; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-49-5009 |
| Medical Device Problem Codes |
Fracture (1260); Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Synovitis (2094); Osteolysis (2377); Inadequate Osseointegration (2646); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a male patient, who had a right knee implanted, underwent a revision procedure approximately 5 years post the initial procedure.The patient presented with increasing knee pain, a sense of instability, and swelling.Plain radiographs revealed polyethylene wear, loosening of both the femoral and tibial components, and osteolysis affecting both bones.During the revision procedure, a femoral and tibial bone defect was identified consistent with aori type 2b.There was also a significant patellar defect and a notable soft-tissue reaction characterized by synovitis.The patient was revised to a competitor¿s devices.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.X-rays were provided.The explanted devices are not available for return.No further information.This is 1 of 4 reports.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6), 02-010-03-0350 - logic femur cr cementado nº5 der.(b)(6), 02-012-45-5040 - bandeja tibial logic fit 5f/4t, (b)(6), 200-02-35 - rotula tres tetones 35mm.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h6, h10, h11.No device was returned for evaluation; the reported event was confirmed by review of radiograph images provided.The most recent pre-revision radiographs showed that the space between the femoral and tibial component appears to decrease between the two provided a-p and sagittal view radiographs, which is a potential sign of insert wear.The femoral component also appears to be closer to the tibial tray near the medial side, which is similar to the wear pattern observed on the insert.In the sagittal view, there appears to be lucencies between the femoral component anterior flange and the femoral bone which indicates femoral loosening.A device image was also provided.There appears to be signs of wear and damage among both the medial and lateral articular surfaces.There are signs of wear and missing material on the posteriomedial aspect of the insert.There is also some observed pitting and signs of delamination and subsurface damage.The full extent of wear cannot be determined due to the presence of biological fluid staining across the explanted insert.No images were provided for the tibial tray, femoral component, or patellar implant.No manufacturing process-related non-conformances, deviations, or exceptions are associated with these devices.A review of complaint history determined that no further action is required as no trends were identified.The revision reported was likely the result of tibial insert prosthesis wear and femoral loosening.Possible causes for polyethylene wear include the confirmed femoral loosening, the alleged joint instability, alleged tibial loosening, high contact stresses during knee flexion, third body wear, patient-related conditions, inclusion of the polyethylene in the packaging recall with the insert being packaged for over five years prior to implantation, or any combination of these possibilities.Instability may be the result of increased soft-tissue laxity (looseness) or the confirmed femoral loosening and subsequent improper alignment and positioning of the prosthesis, but the presence of joint instability cannot be confirmed.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|