| Catalog Number |
SSM-445 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" diagnosed via ultrasound.This record is for the left side.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: ¿ rupture: not observed.As per the investigation procedures, creases and deformation were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "rupture" diagnosed via ultrasound.Device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|