Catalog Number 061.3318
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The removal of a dental implant during surgery without the replacement of another dental implant is a known inherent risk of the procedure due to either lack of primary stability of the implant (patienor procedure related).It may also include the removal of an implant after osseointegration due to either the clinician's or patient¿s decision.The manufacturer¿s trend analysis confirms that usually procedural errors and/or patient's condition contribute to the event.
 
Event or Problem Description
The clinician reports that the day the implant was placed in fdi 16, failure occurred upon insertion.Details of surgery: primary stability not achieved.Patient presented with bone type iv.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLX Ø3.5 RB, SLACTIVE® 18MM, RXD
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key22314811
Report Number0009613348-2025-009069
Device Sequence Number8857426
Product Code DZE
UDI-Device Identifier07630031743329
UDI-Public07630031743329
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191256
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number061.3318
Device Lot NumberKJR86
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/24/2025
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/20/2022