Model Number REF 316-07
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Hypoventilation (1916)
Date of Event 09/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
Subglottic suction line detached from tracheostomy cannula during treatment.
 
Additional Manufacturer Narrative
The investigation of the affected sample was not possible because the product was not returned to trace.Based on the incident description it can be supposed that the defect described relates to the known product error of the subglottic suction line detachment.Further investigations are conducted within capa 000106/330 and follow up.The cause is probably due to the bonding process of the subglottic suction line.It is possible that preparation steps before bonding were not carried out sufficiently or correctly, or that the bonding was not carried out correctly.Other causes could also be seen within the use period, e.G.The use of different cleaning or disinfection agents, humification with a humification mask, as well as strong pulls on the suction line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACOE TWIST PLUS TRACHEOSTOMY CANNULA
Common Device Name
TRACOE TWIST PLUS TRACHEOSTOMY CANNULA WITH CUFF AND SUCTION LINE,
Manufacturer (Section D)
TRACOE MEDICAL GMBH
reichelsheimer str. 1/3
nieder-olm, 55268
GM  55268
Manufacturer (Section G)
TRACOE MEDICAL GMBH
reichelsheimer str. 1/3
nieder-olm, 55268
GM   55268
Manufacturer Contact
margarete halle
reichelsheimer str. 1/3
nieder-olm, 55268
GM   55268
MDR Report Key22317601
Report Number8010485-2025-00028
Device Sequence Number8988333
Product Code JOH
UDI-Device Identifier04035324019076
UDI-Public4035324019076
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K961449
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Remedial Action Replace,Other
Type of Report Initial
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberREF 316-07
Device Catalogue NumberREF 316-07
Device Lot Number1100023432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2025
Initial Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-