BIOTRONIK SE & CO. KG RIVACOR 7 VR-T DF4 PROMRI; ICD Back to Search Results
Model Number 429536
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
This device shows eri with unexpected battery behavior.Device remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
The device was not returned for analysis.The analysis is therefore only based on the returned device data as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was reviewed.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing.The final acceptance test proved the device functions to be as specified.In a next step, the returned device data was inspected.During the inspection the clinical observation could be confirmed.Analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component, which may compromise the ability of the device to deliver therapy.Biotronik has informed the health care professional about this analysis result, who decides, based on the patient¿s individual circumstances and medical judgment, if the device will be exchanged.Should further relevant information or the device itself become available, this investigation will be updated and you will be informed accordingly.
 
Additional Manufacturer Narrative
On (b)(6) 2025 device explanted.The device was not returned for analysis.The analysis is therefore only based on the returned device data as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was reviewed.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing.The final acceptance test proved the device functions to be as specified.In a next step, the returned device data was inspected.During the inspection the clinical observation could be confirmed.Analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component, which may compromise the ability of the device to deliver therapy.Biotronik has informed the health care professional about this analysis result, who decides, based on the patient¿s individual circumstances and medical judgment, if the device will be exchanged.Should further relevant information or the device itself become available, this investigation will be updated and you will be informed accordingly.
 
Additional Manufacturer Narrative
On (b)(6) 2025, device explanted.The icd was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon receipt, the device was interrogated.The interrogation could be properly performed and revealed the mos1 battery status.Battery trend data showed an unexpected voltage drop that resulted in the observed activation of the eri battery status.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a normal and expected sensing and shock delivery.In particular, the specified energy level was reached.At a next step the memory content of the device was inspected.The data showed an inconsistency between the amount of charge taken from the battery and the battery voltage, which led to the automatic activation of the eri battery status.The data analysis confirmed an unexpectedly low battery voltage.In order to obtain further insights, the icd was opened and the inner assembly was inspected.The visual inspection showed no anomalies.Subsequently the current consumption of the electronic module was verified by direct measurement and proved to be as expected and consistent with the interrogated icd data.There was no indication of a malfunction of the electronic module.Therefore, the battery was disconnected from the electronic module and sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The battery was subjected to a visual inspection which did not reveal any signs of damage.In a next step, the battery was opened for destructive analysis.The inspection of the inner assembly identified a damaged insulation, which led to an elevated internal self-depletion.In conclusion, the analysis revealed an elevated internal self-depletion within the battery.Based on the analysis all therapy functions of the icd were entirely available while the device was implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIVACOR 7 VR-T DF4 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22318911
Report Number1028232-2025-03026
Device Sequence Number8858036
Product Code LWS
UDI-Device Identifier04035479156817
UDI-Public(01)04035479156817(17)220131
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number429536
Device Catalogue Number429536
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Supplement Date Received by Manufacturer06/26/2025
Not provided
03/27/2026
Initial Report FDA Received Date06/24/2025
Supplement Report FDA Received Date06/26/2025
06/30/2025
03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age66 YR
Patient SexUnknown
-
-