EXACTECH, INC. OPTETRAK; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
232-03-03 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Osteolysis (2377); Metal Related Pathology (4530)
|
| Date of Event |
03/28/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitants: (b)(6), 21-0901-00t-l1 - stryker s6 90x13/21x1.0, (b)(6), 200-04-33 - (disc) cemented finned t ib.Tra sz 3f/3t, (b)(6), 200-23-09 - cr tibial insert sz3, 9mm.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported a 52 yo male patient who had a right knee implanted, underwent a revision procedure approximately 10 years and 10 months post the initial procedure.The patient was revised due to premature polyethylene wear with massive osteolysis.Associated with a pseudotumoral lesion in the popliteal area secondary to a foreign body reaction caused by polyethylene particles.The patient was revised to a competitor¿s devices.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.X-rays were provided.The explanted devices are not available for return.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h3 - the revision reported was likely caused by tibial insert prosthesis wear over the course of 10 years, which led to metal component wear of the femoral and/or tibial tray.The extent of the wear and possible causes could not be confirmed from the provided information.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation.H6 the following sections were corrected: h6 - codes 4756, 4118, 3233, 11 no longer applies should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|