| Model Number |
2110 |
| Medical Device Problem Codes |
Failure to Calibrate (2440); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: month and year of event have been provided; day is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the cassette sensor encountered an error during testing.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
D9: date returned to mfg; 6/27/2025.Device evaluation: one device received for device analysis.Service history review identified the complaint was not related to a previous service of the device within the review period.Functional testing and visual inspection performed during analysis.The customer reported complaint was confirmed.It was found that the downstream occlusion(dso) sensor was out of calibration.The dso sensor was recalibrated.
|
| |
|