• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 4.0MM FEMALE HEX DRIVER; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 4.0MM FEMALE HEX DRIVER; SCREWDRIVERS Back to Search Results
Catalog Number 210070155
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that broken screwdriver discovered by spd.There was no impact on case.
 
Additional Manufacturer Narrative
J&j medtech is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which j&j medtech has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, j&j medtech or its employees that the report constitutes an admission that the device, j&j medtech, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: investigation summary the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device does not found signs of breakage at the 4.0mm female hex driver.However the sample presents a crack at the hex tip.The potential cause can be attributed to unintended use error, with the damage likely resulting from overtightening, prying motion while device is assembled and use of excessive forces.Proper handling and adherence to the approved use of the device can diminish the risk of failure.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the 4.0mm female hex driver would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0MM FEMALE HEX DRIVER
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22320298
Report Number1818910-2025-10409
Device Sequence Number21063920
Product Code HXX
UDI-Device Identifier10603295085959
UDI-Public(01)10603295085959(10)PG217705
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number210070155
Device Lot NumberPG217705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2025
Initial Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-