• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1650
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the battery does not provide power nor is it charging.The status light of the battery charger is red when the battery is connected to the battery charger.The battery was replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for device evaluation.Product event summary: the battery ((b)(6)) was returned for evaluation.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Visual evidence provided by the site revealed that the status light on a battery charger was flashing red when (b)(6) was connected.Failure analysis revealed that the returned battery passed external visual inspection.Functional testing revealed that the battery flashed red on the battery charger due to a high max error.The high max error is an indicator of how well the battery¿s relative state of charge (rsoc) is calibrated.However, the battery was still able to charge and provide power to a controller.A tendency to exchange batteries before reaching 25% relative state of charge (rsoc) may contribute to an out-of-calibration condition.A high max error will cause a battery to flash red when connected to a battery charger.The high max error was able to be reset and the battery performed as intended.Internal inspection did not reveal any anomalies.Log file analysis revealed no anomalies within the analyzed period.As a result, the reported not charging event could not be confirmed; however, flashing red caused by the high max error condition was likely perceived as the reported not charging event.Based on the available information, the most likely root cause of the reported event can be attributed to a battery that is out of calibration.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22320830
Report Number3007042319-2025-01315
Device Sequence Number21079896
Product Code DSQ
UDI-Device Identifier00763000629717
UDI-Public(01)00763000629717(17)240831
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2024
Device Model Number1650
Device Catalogue Number1650
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Supplement Date Received by Manufacturer08/06/2025
Initial Report FDA Received Date06/24/2025
Supplement Report FDA Received Date08/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1104 VAD
Patient Age75 YR
Patient SexMale
-
-