• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® HEMATURIA COUDE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® HEMATURIA COUDE FOLEY CATHETER Back to Search Results
Catalog Number 2556H20
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that customer called that their bard foley catheter had come out.Showed only a tube and nothing more.A moment ago, there was a pop and fluid was flowing.Takes the bard catheter and takes it out to the sink and checks with colleagues.They conclude that it had broken.It looks like a crack gone through it and they could clearly saw that everything had come out and nothing was left in the patient.Per follow up information received via ibc on 06jun2025, stated that no patient harm.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.Visual evaluation of the returned sample noted one opened (without original packaging), used latex foley catheter.Visual inspection of the sample noted the balloon was ruptured.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: instructions for use the clinical benefits of hematuria/hemostatic hydrogel coated latex foley catheters are that they aid in clinical management by allowing for continuous drainage and measurement of urine as well as bladder irrigation and hemostasis.Hematuria/ hemostatic hydrogel coated latex foley catheters come with a variety of specialized tips used for drainage and clot removal.They have larger balloon sizes and additional lumens to aid in providing bladder irrigation and hemostasis.The hydrogel coating is designed to reduce friction in catheter insertion.Foley catheters can be used in patients of all ages.Up to 10 ch/fr, is generally considered pediatric sizing; however, the appropriate size should be determined by the healthcare professional.Indications for use: hematuria/hemostatic hydrogel coated latex foley catheters are intended for use in the drainage and/or collection and/or measurement of urine.Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.Three-way continuous irrigation foley catheters are indicated for use in bladder/ urinary tract irrigation.Further, foley catheters with larger (30 ml and above) balloons are used to assist in hemostasis following surgery such as transurethral resection of the prostate.Intended user: this product is intended to be used by qualified healthcare professionals.Intended use: for urological use only.Contraindication: no known contraindications.Warnings: on catheter, do not use ointments or lubricants having a petroleum base.They will damage catheter and may cause balloon to burst.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state, and federal laws and regulations.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness, or death of the patient.Users and/or patients within the european union, should report any serious incident that has occurred in relation to the device to the manufacturer and the competent authority of the member state in which the user and/or patient is established.Users outside of the european union should report any serious incident that has occurred in relation to the device to the manufacturer and the regulatory authority of the country in which the user and/or patient is established.Precautions: do not use device if package is opened or damaged.Do not aspirate urine through drainage funnel wall.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Correction: d, e, h this report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that customer called that their bard foley catheter had come out.Showed only a tube and nothing more.A moment ago, there was a pop and fluid was flowing.Takes the bard catheter and takes it out to the sink and checks with colleagues.They conclude that it had broken.It looks like a crack gone through it and they could clearly saw that everything had come out and nothing was left in the patient.Per follow up information received via ibc on 06jun2025, stated that no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® HEMATURIA COUDE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key22320986
Report Number1018233-2025-04953
Device Sequence Number8819555
Product Code KOD
UDI-Device Identifier00801741025112
UDI-Public(01)00801741025112(11)241204(17)290928(10)MYJX4950
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2556H20
Device Lot NumberMYJX4950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2025
Supplement Date Received by Manufacturer08/22/2025
Initial Report FDA Received Date06/24/2025
Supplement Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-