This report is being submitted as part of remediation activities associated with capa 643143, addressing mdr reporting guidance from the fda¿s 1995 preamble, which ensures complaints related to corrections and recalls previously reported to the fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve (b747385), after the valve was reca ptured twice due to deep positioning, the deployment knob of the delivery catheter system (dcs) separated.Hypotension was noted and the handle pieces were quickly reattached, and the valve was recaptured.The system was removed from the patient.The valve became contaminated after removal.A new valve and dcs were used for a successful implant.No adverse patient effects were reported.
|