| Model Number |
8888505016 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the problem was that the tube doesn't suck; it's as if the suction port was blocked by plastic.Per the customer, they used three tubes, and only one worked.There was no harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) for lot number 2506505464 was reviewed and no anomalies related to the reported issue were observed.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported issue.¿ as such, a root cause could not be determined at this time.However, a corrective and preventative action has been opened to address the reported issue.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|