|
The device history record (dhr) for lot number 2506505564 was reviewed and no anomalies related to the reported issue were observed.The physical device was received for evaluation.Upon technical evaluation, it was decided to dissect the returned device to corroborate the reported condition.Only minor marks were observed; no cracks were identified on the returned sample.Functional tests, including a leak test, were performed, and no leaks were detected.Based on these findings, the reported issue has been classified as cosmetic, as it does not impact on the product¿s form or function.Based on this information, no corrective actions are warranted at this time.We will continue to track and trend through our monitoring programs and take actions as applicable.
|