U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. DISPOSABLE BIOPSY FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. DISPOSABLE BIOPSY FORCEPS Back to Search Results
Model Number FB-231D
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/28/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is linked to the following patient identifiers: (b)(6).The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that during a diagnostic endobronchial ultrasound (ebus) procedure; while using the broncho videoscope along with a single-use biopsy valve, a piece of the biopsy valve fell off into the patient¿s body when the biopsy forceps were inserted into the scope and reached the distal end of the scope.After the procedure, inspection of the biopsy valve revealed that the seal ring inside the biopsy valve was chipped.It is suspected that the seal ring may have fallen into the patient¿s body.While attempting to retrieve the fallen object, the patient began coughing, and the object was lost from view.The fallen object was not retrieved.It was unclear whether the object remained in the lung or the stomach.According to the physician, it is likely that the object was expelled toward the stomach as a result of the coughing.A lung examination was performed, but the fallen object could not be found.Additional information was received confirming that diagnostic imaging was used and the patient was under sedation during the procedure.The procedure was completed using the same device with a delay of more than 30 minutes.There were no further reports of patient harm.
 
Event or Problem Description
No additional information received from customer.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.It is likely that the slit in the rubber valve or the area around the hole was damaged by overloading due to insertion and withdrawal of the biopsy forceps at an angle to the biopsy valve.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE BIOPSY FORCEPS
Common Device Name
DISPOSABLE BIOPSY FORCEPS
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22324205
Report Number9614641-2025-01053
Device Sequence Number8794126
Product Code EOQ
UDI-Device Identifier04953170158209
UDI-Public04953170158209
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K950636
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFB-231D
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2025
Supplement Date Received by Manufacturer08/03/2025
Initial Report FDA Received Date06/24/2025
Supplement Report FDA Received Date08/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BF-P290 BRONCHOVIDEOSCOPE.; MAJ-210 SINGLE USE BIOPSY .
Patient SexUnknown
-
-