• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090W
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the programmer demonstrated the autonomous cursor movement.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer's comment that the programmer demonstrated the autonomous cursor movement.It was observed that the programmer's finger touch was functioning properly.Additionally, it was noted that the printer drawer release lever and the power cord bay latch tabs were broken off, and the keyboard rail cover was cracked.All found defective parts were replaced, and all other identified issues were resolved.Reconfigured hard drive, reloaded and updated software, and the programmer then passed all final functional and systems tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22325472
Report Number2182208-2025-03038
Device Sequence Number8988582
Product Code KRG
UDI-Device Identifier00885074301440
UDI-Public00885074301440
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2090W
Device Catalogue Number2090W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2025
Supplement Date Received by Manufacturer07/08/2025
Initial Report FDA Received Date06/25/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/30/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-