| Model Number |
VTICM5_12.6 (-10.50/1.50) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a lens rotation not associated to a low vault.The lens was repositioned but that did not resolve the problem.On a later visit the lens was exchanged with a spherical lens of the same length but different power.In a separate visit an additional lri was performed for astigmatism and the problem resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition is identified in the labeling as a known potential adverse event following icl implantation.H3: device evaluation is required but has not yet been performed.H6: investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned dry with residue/debris on the lens in a microcentrifuge vial.Visual inspection found no visible damage to the lens.Dimensional inspections found the lens to be within specifications.Claim: (b)(4).
|
| |
|