This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting guidance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Unit received with unexpected battery power loss and had high sleep current, problem isolated to pcb 2.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported to medtronic minimed that the customer experienced frequent replace battery now alarms, along with rapid battery depletion within 3-4 days, and issues with sensor connectivity.The customer reported no adverse event.The event involved product(s) mmt-7020a and mmt-1780kpk.Troubleshooting was performed for both battery and sensor issues, but the customer continued to experience the same problems.No harm requiring medical intervention was reported.The products (pump and sensors) were replaced, and the pump was returned for analysis.
|