| Catalog Number |
RBYPOD8 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the splenic artery using pod coils, a pod packing coil (packing coil), and a non-penumbra microcatheter.During the procedure, while advancing a pod coil through the microcatheter, the pod coil unintentionally detached within the microcatheter.Therefore, the microcatheter containing the detached pod coil was removed.The same issue occurred with another pod coil and a packing coil.The physician removed the microcatheter containing the detached pod coil and packing coil respectively.The procedure was completed using a new microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the following section was updated on this follow-up #01 mfr report 3005168196-2025-00213: 1.Section b.Box 5.Describe event or problem.Evaluation of the first returned pod coil could not confirm that the embolization coil was detached from its pusher assembly.Evaluation revealed that the pet lock was intact, the embolization coil was intact with the pusher assembly and undamaged.Therefore, the root cause of the reported complaint could not be determined.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.Note: based on the investigation findings, this is not considered a reportable event.This event did not and, if it were to recur, it would not cause or contribute to serious deterioration or death.This report is also associated with mfr report number: 1.3005168196-2025-00215.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the splenic artery using a pod coil and a non-penumbra microcatheter.During the procedure, while advancing a pod coil through the microcatheter, the pod coil unintentionally detached within the microcatheter.Therefore, the microcatheter containing the detached pod coil was removed.It was reported that the physician was unable to implant a pod coil and a pod packing coil (packing coil) and no additional details are available.Therefore, both coils were removed.The procedure was completed using a new microcatheter.There was no report of an adverse effect to the patient.
|
| |
|