| Model Number |
A22042A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.It was observed that during the device inspection the resection sheath, 26 fr, the ceramic tip was broken.A root cause was determined to be degradation leading to component failure should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the resection sheath, 26 fr, the ceramic tip was broken.There was no patient involvement.
|
| |
|