• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number IT2110X21B
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo ventilator was returned to a third-party service center for required maintenance.There was no patient involvement, and no injury or harm was reported.During the evaluation of the device at the manufacturer's service center, error codes related to the internal battery were confirmed in the ventilator's downloaded error log.The internal li-ion battery required replacement to address the issue.Additional findings not related to the malfunction/issue: the detachable battery was found to be defective and was replaced.The system printed circuit board assembly was replaced due to an error code (e-290) indicting a start/stop key latch failure.As part of maintenance the blower, low flow oxygen tubing, machine flow sensor, a proximal in-line filter and the in-line filter were all replaced for contamination.Four-year maintenance service was completed with filter replacements per maintenance schedule.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key22326841
Report Number2518422-2025-106828
Device Sequence Number21203104
Product Code CBK
UDI-Device Identifier00606959055421
UDI-Public(01)00606959055421
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIT2110X21B
Device Catalogue NumberIT2110X21B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2025
Initial Report FDA Received Date06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/03/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-