Catalog Number 21-2966-24
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event or Problem Description
It was reported that the device exhibited a leakage at the collar port near the access needle.There was patient involvement, no patient injury was reported.
 
Additional Manufacturer Narrative
The device history record of reported lot number 6023345 was reviewed, and it was confirmed that no non-conformities were reported during production.The quality inspection forms were reviewed, and it was observed that no defects were found, and the lot was released.Without the return of the samples a comprehensive failure investigation cannot be performed, and a probable cause cannot be determined.If the product is returned, the complaint will be reopened for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIPPER PLUS SAFETY PORT ACCESS NEEDLE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key22327545
Report Number3012307300-2025-07622
Device Sequence Number18214960
Product Code FPA
UDI-Device Identifier10610586027741
UDI-Public(01)10610586027741(17)290607(10)6023345
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number21-2966-24
Device Lot Number6023345
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date06/25/2025
Supplement Report FDA Received Date08/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexFemale
-
-