Foreign object found in keytruda bag.Event details: keytruda was prepared using two vials, and the drug solution was collected from each vial and injected into an infusion bag.When checked the bag before dispensing the infusion bag to the nurse, a very small foreign object was found.We would like an analysis of the size of the coring.(contaminated in the bag) the injector's lot is unknown.The incident occurred after the protector and injector were connected.Additional information: the product that was connected to the infusion bag was 515309 1: this is the same injector.515005 lot is unknown.2: the l connector is connected, and the collected medication is injected from the l connector into the infusion bag.
|
|
(b)(4) follow up mdr for device evaluation: it was reported that particles were observed inside an iv bag.For investigation, two vials, one iv bag, one l-connector, and a device not manufactured by bd were provided to the quality team.Upon inspection, a gray particle was confirmed inside the infusion bag.Testing determined that the particle was consistent with material from the infusion bag's rubber stopper.No damage or defects were identified within our device that could have contributed to this observation.Additionally, particles were noted within the vial itself, which also appear to be consistent with material from the vial stopper.Coring of a rubber stopper can occur depending on stopper quality, stopper design, or if an incomplete or misaligned connection is made during spiking.Finished products for this reference undergo sampling as well as visual and functional inspections throughout the various manufacturing subprocesses in accordance with established procedures.Because a lot number was not provided for this incident, a device history record review could not be completed.Based on the available information, we cannot definitively determine which spike was responsible for introducing the particle observed in the iv bag; therefore, a root cause cannot be established at this time.The quality team will continue to monitor complaints for this device and reported condition for any emerging trends.
|