| Catalog Number |
20-750 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: rupture; capsular contracture, baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade unknown, "lower", and "rupture"; ultrasound performed.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, corrected and/or changed data: b.5, d.6b, d.9, h.3, h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported capsular contracture baker grade unknown, "lower" and "rupture".This record is for the right side.The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade unknown, "lower", and "rupture"; ultrasound performed.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: rupture: not observed.Capsular contracture: unable to observe as it is not related to the manufacturing process.A workmanship was observed not related to the complaint for a split in patch event.A further investigation is requested.As per the investigation procedure, deformation and crease were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Further investigation: the review of the documentation associated to the manufacturing process indicates that all devices involved in this work order were released in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents or in the personnel training related to the reported event.Additionally, no er/ ncr(s) were identified during the investigation associated with this lot and the complaint.According to the device analysis performed in the laboratory was observed split (ss) in patch area, it is out of specification per qa 199.04.The events of rupture was not observed and capsular contracture was unable to observe, and the workmanship has not relation to reported complaint.Therefore, the reported event was not confirmed.A review of the current risk documents pfmea-silicone filled bi and sizer assy rev 8.0 was performed, and the event of split in patch is a known event, for which current process controls and inspections are established to reduce the incidence of this defect.Considering that there is not an adverse trend for this type of event only inv-73303 in (b)(6) 2023 no additional actions are deemed required at this time.The issues with split in patch will continue to be monitored and corrective action taken in the future if deemed appropriate.Additional, changed, and/or corrected data: h3, h6, h11.
|
| |
|