| Catalog Number |
14-534900 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that both prongs on the tip of a zyston inserter broke off while inserting the cage into the disc space intra-operatively; the connection between the inserter and cage loosened.The threads of the cage were damaged while the surgeon was backing it out to exchange inserters; the cage was also replaced to complete the procedure without patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: the remainder of the udi number is unknown because the lot number is not available.D10: 14-534187 lot 0100371 (zyston straight 7h x 20lx10w lordotic spacer).Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: this event could possibly be attributed to off-axis or excessive forces applied during use.Dhr review: the lot number was not reported and the device was not returned, therefore a dhr review is unable to be performed.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that both prongs on the tip of a zyston inserter broke off while inserting the cage into the disc space intra-operatively; the connection between the inserter and cage loosened.The threads of the cage were damaged while the surgeon was backing it out to exchange inserters; the cage was also replaced to complete the procedure without patient impacts.
|
| |
|