Model Number IT2110X21B
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilator service required alarm occurred along with a request for preventative maintenance.The device was reported to be in patient use.There was no allegation of harm or injury reported.During the evaluation of the device at the third-party 's service center error codes related to failure of the internal li-ion battery was observed.The technician recommended replacing the device's internal battery to address the issue.No further investigation is required at this time.Additionally, the device's muffler assembly, air foam inlet filter, and particulate filter will be replaced according to the 4 year preventive maintenance procedure.The device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO ITALY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22328464
Report Number2518422-2025-106836
Device Sequence Number8300520
Product Code CBK
UDI-Device Identifier00606959055421
UDI-Public(01)00606959055421
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIT2110X21B
Device Catalogue NumberIT2110X21B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Initial Report FDA Received Date06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-