(b)(4).The customer returned one 18 ga introducer needle for analysis.Signs of use were observed on the returned component.Visual analysis revealed that the needle hub was cracked into pieces.The missing pieces were not returned.The outer diameter of the needle cannula measured 0.0495", which is within the specification limits of 0.0495" - 0.0505" per cannula product drawing.The inner diameter of the needle cannula measured 0.041", which is within the specification limits of 0.041" - 0.043" per cannula product drawing.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user , "some disinfectants used at catheter insertion site contain solvents which can weaken the catheter material.Alcohol, acetone, and polyethylene glycol can weaken the structure of polyurethane materials.These agents may also weaken the adhesive bond between catheter stabilization device and skin." the customer report of needle hub cracked was confirmed by visual inspection of the returned sample.Visual analysis revealed that the needle hub was cracked into pieces.The failure mode identified is consistent with the failure mode investigated per a previously opened capa.Based on the capa investigation, the root cause of this complaint is design-related.Teleflex has identified that this needle hub material is susceptible to cracking when placed under stress (i.E.Pressed onto a tapered luer fitting, sideloaded as the clinician attempts to locate a vessel, etc.) in the presence of liquid alcohol-based disinfectants.The capa was implemented using a new alcohol-resistant material to manufacture the needle hub.The needle hub from this complaint was confirmed to be made from the old needle hub material.Teleflex will continue to monitor and trend for reports of this nature.
|