| Model Number |
25-473-80-07 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported rpms exceeded the maximum recommended speed found in the ifu.
|
| |
|
Event or Problem Description
|
|
It was reported during initial surgery a twist drill fractured during use.A portion of it remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed.The results of the investigation have identified no additional actions are necessary at this time.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|