• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE I/A HANDPIECE TIP; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE I/A HANDPIECE TIP; CATHETER, IRRIGATION Back to Search Results
Catalog Number 8065752066
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that the phacoemulsification (phaco) tip was found bent during the calibration step of surgery.The type of surgery was unknown.The surgery was completed by replacing the product.There was no report of patient impact.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.One opened phacoemulsification (phaco) tip, in a wrench was received.Sample was visually inspected and found to be nonconforming with phaco was bent at cone to cannula transition area.Wear observed on threads, back of flange, or nut corners, consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Attached photo was reviewed by the manufacturing site.The photo shows two phaco tips in a wrench in a blister.Reported issue cannot be confirmed from photo.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.The returned sample is visually non-conforming with the phaco tip bent; therefore, a bent phaco tip was confirmed; however, how and when the phaco tip became bent could not be determined.Most likely root cause is handling during placement of the phaco tip onto the handpiece.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I/A HANDPIECE TIP
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22329482
Report Number2523835-2025-00702
Device Sequence Number8301044
Product Code GBX
UDI-Device Identifier00380657520664
UDI-Public(01)00380657520664(17)270831(11)220908(10)1511AK
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910245
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752066
Device Lot Number1511AK
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/23/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/04/2025
Supplement Date Received by Manufacturer09/01/2025
Initial Report FDA Received Date06/25/2025
Supplement Report FDA Received Date09/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-