| Catalog Number |
365724 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that for 15 years, the user had no issues using a green foley catheter from the company.Recently, they received a brown replacement due to supply issues.After one day, the user experienced urine and fluid leakage and informed their medical team.The catheter was difficult and painful to remove, with small bubbles noted at the end.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The reported product number is unknown.The udi number for this product is not available.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It was unknown whether the device had met relevant specifications.The potential root cause for this failure could be user related (example: contact with sharp object)/ mechanical failure/ operator error/ thin rubberize layer.A dhr review is not required as the lot number is unknown.Unable to review the labelling due to unknown product code.Although the product family is unknown, the foley catheter ifus are found to be adequate based on past reviews.Correction- e.The reported product number is unknown.The udi number for this product is not available.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that for 15 years, the user had no issues using a green foley catheter from the company.Recently, they received a brown replacement due to supply issues.After one day, the user experienced urine and fluid leakage and informed their medical team.The catheter was difficult and painful to remove, with small bubbles noted at the end.Per additional information received via email on 10jul2025, customer stated they are maintaining the catheter currently in a ziplock bag.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed because the reported failure could not be reproduced.No root cause could be found because the reported event was unconfirmed.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and states the following: insert foley catheters only for appropriate indications and leave in place only as long as needed.Proper techniques for urinary catheter insertion: - perform hand hygiene immediately before and after insertion - insert urinary catheters using aseptic technique and sterile equipment - use the smallest foley catheter possible, consistent with good drainage - document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record proper techniques for urinary catheter maintenance: - secure the foley catheter.Use the statlock foley stabilization device if provided - maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions - maintain unobstructed urine flow and keep the catheter and collection tube free from kinking - keep the collection bag below the level of the bladder or hips at all times - empty the collection bag regularly using a separate, clean collection container for each patient.Correction- d, g, h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that for 15 years, the user had no issues using a green foley catheter from the company.Recently, they received a brown replacement due to supply issues.After one day, the user experienced urine and fluid leakage and informed their medical team.The catheter was difficult and painful to remove, with small bubbles noted at the end.Per additional information received via email on 10jul2025, customer stated they are maintaining the catheter currently in a ziplock bag.
|
| |
|