• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON PHASEAL; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON PHASEAL; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Catalog Number 515106
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
Foreign object in bag.Event details: prepared 1 vial of 100 ml paclitaxel and 1 vial of 30 ml paclitaxel, then checked the bag and found a single foreign object.Eight vials of paclitaxel nk 30 ml were prepared, and after collecting the liquid from each vial, it was infused into the infusion bag.Before the infusion bag was handed over to the nurse, the person in charge checked the bag and found multiple foreign objects.The lot number of the injector is unknown.The incident occurred after connecting the protector and injector.The connection was made approximately 10 times.The product that was connected to the infusion bag was 515309.
 
Additional Manufacturer Narrative
Follow up mdr for device evaluation/correction: based on investigation results, this complaint is no longer reportable.The involved device did not contribute to the reported failure, coring.It was reported that particles were observed inside the iv bag.For investigation, two iv bags, two l-connectors, and a device not manufactured by bd were provided to our quality team.Upon inspection, gray particles were observed inside both infusion bags, verifying the reported concern.The particles were evaluated under a microscope, and it was determined that they are consistent with material from the rubber stopper of the infusion bag.No damage or related defects were identified within our device that could have contributed to this observation.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.As a lot number was unavailable for this incident, a device history record review could not be completed.Coring of the rubber stopper may occur depending on stopper quality, design, or if a poor connection is made.Based on the available information, a definitive root cause cannot be identified at this time.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHASEAL
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22329829
Report Number3003152976-2025-00324
Device Sequence Number9354854
Product Code ONB
UDI-Device Identifier00382905151064
UDI-Public(01)00382905151064
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number515106
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2025
Supplement Date Received by Manufacturer01/14/2026
Initial Report FDA Received Date06/25/2025
Supplement Report FDA Received Date01/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-