| Model Number |
NA-U401SX-4021 |
| Medical Device Problem Codes |
Material Too Rigid or Stiff (1544); Failure to Fire (2610); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: related report number: report 2: 3011050570-2025-00602.Related patient identifiers: report 3: (b)(6), report 4: (b)(6).The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the subject device handle mechanism was overly stiff which caused problems deploying the needle properly and caused the needle tip to bend.This event occurred during a diagnostic endobronchial ultrasound procedure.The procedure was completed with a similar olympus device.There were no reports of patient harm.This is report one of four.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: h2, h4, h6, h11.The device was not returned for evaluation and the customer's allegation could not be confirmed.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|