| Catalog Number |
305892 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd needle eclipse leaked at connection with syringe.The following information was provided by the initial reporter: on (b)(6) 2025 vaccine/liquid has leaked on administration of menquadfi vaccination, needle did not become detached.Parent has been contacted and vaccination will need to be repeated.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was completed for provided material number 305892 and batch number 2411007.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported defect and all quality tests were found to be within specification.As samples were unavailable for return, twenty (20) retained samples of reported lot number were obtained from the manufacturing facility.The needles were assembled with a bd discardit 2ml syringe; however, no signs of defective hub connection were identified nor were any signs of leakage detected.Based on the investigation results, we were unable to reproduce the reported issue and an exact cause could not be determined.Engagement force is measured during manufacture as part of the in-process inspections and final testing prior to release.Smartslip technology has been introduced to ensure that a secure connection is made with luer slip syringes.We recommend following the instructions for use, which are unique in directing the user to push firmly when attaching the needle to the syringe and to be sure that the clip is activated.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|