| Catalog Number |
68LP-440 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the right side.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed, one opening assessed as fold flaw opening.As per the investigation procedure, crease and wear abrasion were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the right side.The device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|