| Catalog Number |
0279250101 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that parts of the tip broke off during procedure.All of the broken pieces were retrieved from patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that parts of the tip broke off during procedure.All of the broken pieces were retrieved from patient.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the device will not be returned for investigation.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip broken probable root cause: non-conforming component, poor assembly process, misuse.Use errors probe used as a tool for mechanical displacement of tissue or bone, or to pry or scrub excessive force used when inserting of removing probe, probe inserted into obstructed passageway, or any forceful impact to the tip of the probe with rigid objects excessive use of device beyond stated lifetime the reported failure mode will be monitored for future reoccurrence.
|
| |
|