| Model Number |
G28783 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, prior to use for an unknown procedure in interventional radiology, foreign matter resembling a hair was noted in an amplatz stiff support wire guide's sealed sterile package.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
E3: occupation - sr.Distribution specialist.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Additional information was received 05aug2025.The package was not opened, and the device was not used.The procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary of event: as reported, prior to use for an unknown procedure in interventional radiology, foreign matter resembling a hair was noted in an amplatz stiff support wire guide's sealed sterile package.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.Additional information was received 05aug2025.The package was not opened, and the device was not used.The procedure was completed successfully.Corrected information: d4 = udi.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), specifications, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The sealed/unused complaint device was returned to cook for investigation.A hair-like fiber was noted in the pouch.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the lot.A review of complaint history found no additional complaints for this lot number.The product ifu states ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification; however, there is no evidence of additional non-conforming devices in-house or in the field.There are 100% inspections in place to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Based on the information provided and the results of the investigation, cook has concluded that a manufacturing and quality control deficiency contributed to this event.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|