• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG L12 LED LIGHT SOURCE WITH AIM; CONFOCAL OPTICAL IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG L12 LED LIGHT SOURCE WITH AIM; CONFOCAL OPTICAL IMAGING Back to Search Results
Catalog Number 0220240300
Medical Device Problem Codes Electrical Power Problem (2925); Optical Problem (3001); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the installed light source has a faulty electrical connection that turns itself on and off.Therefore, there was loss of light.
 
Additional Manufacturer Narrative
Three attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.No further attempts to retrieve the device are required.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: turns itself on and off probable root cause: receptacle board leaf spring poor contact to contact ring, esst voltage malfunction, esd shock to contact ring, poor alignment of cable in jaw, assembly/grounding, light engine malfunction, main board malfunction, receptacle board malfunction, front board malfunction, damaged/missing esst spring, communication with ccu, overheating, power supply malfunction, software malfunction, hf/rf interference - device settings - uc1 on main board - uc2 on main board - uc3 on main board - receptacle board - ccu usb ports - hub port - service port - device controls - uc1 programming port - uc2 programming port - uc3 programming port - receptacle board prog.Port - light engine - debug device functionality - physical location of l12 - l12 firmware operating system the reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the installed light source has a faulty electrical connection that turns itself on and off.Therefore, there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG L12 LED LIGHT SOURCE WITH AIM
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andy dobos
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key22332756
Report Number0002936485-2025-00585
Device Sequence Number8994288
Product Code OWN
UDI-Device Identifier07613327572124
UDI-Public07613327572124
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K230754
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0220240300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2025
Supplement Date Received by Manufacturer08/12/2025
Initial Report FDA Received Date06/25/2025
Supplement Report FDA Received Date09/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-