B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; Implant, endosseous, root-form Back to Search Results
Model Number IRTCA3310
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician indicates that he placed the implant on (b)(6) 2024- and 4-months later it was noted that the implant had not osseointegrated.Patient with bone quality iii.In the event the patient experienced: bone loss.
 
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
Implant, endosseous, root-form
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
Manufacturer Contact
paula sanchez
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
MDR Report Key22335013
Report Number3004417597-2025-00091
Device Sequence Number9024650
Product Code DZE
UDI-Device Identifier08435389812821
UDI-Public08435389812821
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K151391
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberIRTCA3310
Device Lot NumberBD00055.A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/30/2025
Initial Report FDA Received Date06/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
-
-