• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SIZE 4 TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SIZE 4 TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33650004
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Osteolysis (2377); Inadequate Osseointegration (2646)
Date of Event 09/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received a report from the national joint registry that contains unpublished collected data on the usage and the outcomes of the ankle joint replacement.The report details analysis provided for revision procedures performed until (b)(6) 2024.During the review of the report, it was identified that on a patient required revision surgery due to aseptic loosening of tibial component, aseptic loosening of talar component, lysis of tibia, lysis of talus, which was not previously reported to the manufacturer.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation, and no other additional information was received from the national joint registry.More detailed information about the patient's medical history, the event details and the involved device(s) must be available to determine the root cause.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.Please note, that reports received from the national joint registry are not published reports and therefore web link is not available.
 
Event or Problem Description
The manufacturer received a report from the national joint registry that contains unpublished collected data on the usage and the outcomes of the ankle joint replacement.The report details analysis provided for revision procedures performed until (b)(6) 2024.During the review of the report, it was identified that on a patient required revision surgery due to aseptic loosening of tibial component, aseptic loosening of talar component, lysis of tibia, lysis of talus, which was not previously reported to the manufacturer.
 
Additional Manufacturer Narrative
Correction - please refer h6 device code.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY TIBIAL TRAY SIZE 4 TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22335706
Report Number3010667733-2025-00777
Device Sequence Number8962894
Product Code HSN
UDI-Device Identifier00840420122513
UDI-Public(01)00840420122513(11)190910(17)270908(10)1663100
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33650004
Device Lot Number1663100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/05/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date06/26/2025
Supplement Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexUnknown
-
-