|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: e1 initial reporter facility name: (b)(6) medical center on behalf of (b)(6) center.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10 (concomitant) h3, h6: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history review (dhr): a review of the receiving inspection (ri) for ao handle torque limiting3.0nm, was conducted identifying that lot number km881266 was released in four batches.Batch 1: lot qty of (b)(4) units were released on apr 8, 2020 with no discrepancies.Batch 2: lot qty of (b)(4) units were released on apr 6, 2020 with no discrepancies.Batch 3: lot qty of (b)(4) units were released on apr 18, 2020 with no discrepancies.Batch 4: lot qty of (b)(4) units were released on apr 18, 2020 with no discrepancies.Supplier: (b)(4).The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|