• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME; MOTOR, DRILL, ELECTRIC Back to Search Results
Catalog Number CRANI-A-G1
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the adult cranitome device had an undetermined malfunction.During service and evaluation, it was determined that the device produced heat, had component damage and the locking mechanism was stiff/stuck.It was further determined that the device failed pretest for visual assessment and lock operation.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to user, which is user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT CRANIOTOME
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 
3035526892
MDR Report Key22336473
Report Number1045834-2025-00396
Device Sequence Number13051533
Product Code HBC
UDI-Device Identifier00845384016410
UDI-Public(01)00845384016410
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCRANI-A-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/17/2025
Initial Report FDA Received Date06/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-