The device was returned to olympus for inspection.Based on the results of the investigation and previous investigations through the product technical investigation (pti) process, reasonably known information suggests probable causes such as damage or deterioration of parts, environment problem like as dust/dirt/humidity, optical problem, overheating problem, and electrical problems like as signal loss or open circuit.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|